Figure 2: Failure of percutaneous retrieval: (a and b) AP projection fluoroscopy, device in abdominal aorta at T12 level, (c) AP projection with contrast, PA end of occluder stuck at superior mesenteric artery origin without any vascular compromise, (d) 5F pigtail catheter from right femoral artery, goose-neck snare through a 6F multipurpose catheter from IVC–>RA–>RV–>PA–>PDA–>DTA, trying to snare the device with Pigtail catheter supporting the device, (e) Pigtail catheter holding the device in the loop, Maslanka® grasping forceps (BVM Medical Limited, Leicestershire, UK) trying to catch the device, (f) 5F Guide catheter from femoral artery used to rotate the device and trying to snare

Figure 2: Failure of percutaneous retrieval: (a and b) AP projection fluoroscopy, device in abdominal aorta at T12 level, (c) AP projection with contrast, PA end of occluder stuck at superior mesenteric artery origin without any vascular compromise, (d) 5F pigtail catheter from right femoral artery, goose-neck snare through a 6F multipurpose catheter from IVC–>RA–>RV–>PA–>PDA–>DTA, trying to snare the device with Pigtail catheter supporting the device, (e) Pigtail catheter holding the device in the loop, Maslanka<sup>®</sup> grasping forceps (BVM Medical Limited, Leicestershire, UK) trying to catch the device, (f) 5F Guide catheter from femoral artery used to rotate the device and trying to snare