Go to: We report a case of an atrial septal defect where a Figulla Occlutech device embolized into the right ventricle. As this device has no left atrial hub, we brought the bare device into the inferior vena cava and used a novel two-snare technique to slenderize the device into the sheath. This report highlights additional tips in transcatheter device retrieval.


How to cite this article:
Sadiq M, Rehman AU, Mehmood A. Successful retrieval of a Figulla Occlutech septal occluder - embolized device stability and potential solutions.Ann Pediatr Card 2016;9:161-163


How to cite this URL:
Sadiq M, Rehman AU, Mehmood A. Successful retrieval of a Figulla Occlutech septal occluder - embolized device stability and potential solutions. Ann Pediatr Card [serial online] 2016 [cited 2021 Nov 28 ];9:161-163